Brand PLT™ Pipette Leak Testing Unit |
---|
◈
제품정보
The
most frequent cause of inaccuracy in piston-operated pipettes is leakage. This
arises from damage either to the seals, pistons, or tip cones.Often not
detectable by the naked eye, leaks lead to significant volume errors. The BRAND
pipette leak tester (PLT unit) for air displacement pipettes finds even the
smallest leaks within seconds.
Air-displacement pipettes must be checked
at regular intervals and the results must be compared with the ISO 8655-2 error
limits. However, a calibration certificate only reflects the results at the time
of testing. The time between these calibrations is crucial, since leaks can
occur at any time. Well over 80% of pipettes sent in for repair have leaks and
are outside their volume tolerances, even if they don’t drip.
While the
PLT unit cannot replace regular gravimetric testing, daily pipette checks can
provide a safeguard during the periods between calibrations. Even the smallest
leaks are detected! Process reliability for the pipettes is thus significantly
improved.
Leak rates, their detection and quantification
The
leak rate is a measure of the quantity of material that flows through a leak per
unit time. For air-displacement pipettes the PLT unit determines the rate
through a differential pressure measurement, i.e., after creating a negative
pressure, the pressure rise over a given time is
measured.
·
The
leak rate is determined by considering complex physical relationships.
Calculation of the limit values resident in the PLT must include factors such as
the dead volume of the pipette/tip system, flow cross section of the pipette
tip, pressure rise per unit time, pipette volume and type,
etc.
·
The
pV value is the product of the pressure and the volume of a certain quantity of
a gas at the prevailing temperature. This is a measure of the quantity of
material or the mass of the gas.
·
The
leak rate QLis the ratio of the pV value and the period of time
during which the gas flows through a
path
cross-section.
·
For
the pipette test, hPa ml/s is a suitable unit for the leak rate. A leak rate of
e.g., 1 hPa ml/s at an air pressure of 1000 hPa (approx.atmospheric pressure)
means a volume loss of about 1 μl/s.
Easy
To Use
Menu-driven
software, combined with a jog wheel and affirmative/negative buttons make it
easy to select the correct test parameters, pressure units or language (English,
Spanish or German)
A Diagnostic Tool
To check the overall
pipette system, the test is conducted with mounted, unused tip. When a leak has
been identified, the test can be repeated without a tip to determine whether the
leak arises from the tip cone/tip coupling region. Additionally static or
dynamic (piston operated during test procedure) testing can also help localize
source of leaks when detected. The ability to test multichannel pipettes four
channels at a time, or individually increases the throughput of testing, and
assisting in the diagnosis of faulty channels.
Pass/Fail
Determination
The limit values referenced during testing represent a
warning limit, from which significantly lower volume values can also be
determined gravimetrically. This is one quarter of the volume tolerances,
according to ISO 8655-2.
The limit value for the leak volume of a given
pipette allows the leak rate to be calculated. These calculations, which are
based on over 35 years of experience in the development and production of
pipettes, include the dead volume and the intake characteristics, among other
things. If the pipette is mechanically defect-free, clean, and the test is
carried out properly with the BRAND PLT unit, then the instrument is within the
ISO 8655-2 tolerances.
◈
주문정보
Part
No. |
Description |
2703970 |
BRAND
PLTTM Pipette Leak Testing Unit, with universal power
supply |
Accessories
703977
: Multichannel
pipette adapter for BRAND PLT unit. Inlcudes four plugs 703975
: Single
channel pipette adapter for PLT unit, for testing with tips. Includes
plug 703976
: Single
channel pipette adapter for PLT unit, for testing without tips. Includes
plug 703978
: Replacement
polyethylene filters for BRAND PLT unit. Pack of 10 703979
: Universal
power supply for BRAND PLT unit, 100-240V, 50/60Hz |