LSP 주사기펌프는 설치와 작동이 간편하고 높은 정밀도와 신뢰성을 제공하는 실험실용 표준형 주사기펌프이다.
주입전용 제품으로 공급되며, 10 µL에서 60 mL 주사기를 장착할 수 있다.
대형 LCD 화면에서 회전스위치와 멤브레인키패드를 이용하여 파라메터의 설정을 간편하게 할 수 있다
특 징
Syringe Selection : The syringe can be selected in the manufacturer table which includes manufacturer, material and size or input the inner diameter of the syringe barrel directly
Easy to operate: Combining big screen LCD display with rotary coded switch and membrane keypad makes the operation simple and prompt.
Working Mode: Infusion
Memory Function : 1. The parameters are saved in EEPROM. The parameters don't need to be reset when power returns after an interruption 2. In flow rate mode, the pump remains running or stop according to the setting parameters when power returns after an interruption
Protection Function : The pump will stall and give an alarm when the drive structure of the pump is blocked
Communication Function : Realize computer control through RS485 communication interface
Calibration Function : Acquire accurate volume through calibration: Input/output control
Syringe Protection : Adjust syringe rest to prevent syringe from damaging
규 격
확대한 그림보기->
LSP01-1A
LSP01-2A
LSP02-2A
LSP04-1A
Working Mode
Infusion
Channel
1
2
4
Syringe
10 uL ~ 60 mL
10 uL ~ 10 mL
Stroke
140 mm
Pusher Advance
0.156 um/step
0.03125 um/step
0.156 um/step
Linear Speed
5um/min~65 mm/min
1 um/min ~ 13 mm/min
5um/min~65mm/min
Linear Resolution
5 um/min
1 um/min
5 um/min
Linear Accuracy
≤±0.5% when ≥30% of pump stroke
Linear Force
≥9 kg (20 lbs)
≥29 kg (64 lbs)
≥18 kg (40 lbs)
Syringe Selection
Built-in syringe branched, size and inner diameters
Syringe Defined
Can store four user-defined syringe inner diamerters
Flowrate Calibration
Improve flowrate accuracy
Running Parameter
Infusion volume, infusion time, etc
Display Setting
Display volume, flowrate or linear speed
Power-off Memory
Storing the running parameters automatically
State Output
OC gate signal to indicate start/stop and direction
The flowrate can be calibrated for more accurate dispense volume.
1. In parameters setting interface, press and turn RES to highlight the Calibration line. Press RES to enter calibration interface. 2. Testing time line is highlighted. Press RES to highlight the numerical value position. Turn RES to select the testing time. The testing time is from 0.5 minutes to 60 hours. Press RES to confirm. 3. Press Run/Pause key to start the testing. After testing, turn RES to highlight the testing volume line. Press RES to highlight the numerical value position. Enter the actual volume from the membrane keypad. Press RES to confirm. 4. After calibration, the linear speed of the Pusher Block changes to increase the accuracy of the dispense volume.
LSP01(04)-1A Syringe Pump Power Up (Release Time:2017-01-06)
This option is only applicable when no dispense volume is selected. When power returns after an interruption the pump can resume operation (select Run) or remain stopped (select Stop).
LSP01(04)-1A Single Channel Syringe Pump Flowrate setting (Release Time:2017-01-06)
In parameters setting interface, turn and press RES to highlight the flowrate setting line. Press RES to make the numerical value position highlighted. Enter the flowrate required from the membrane keypad. Press RES to set flowrate unit. Turn RES to scroll through the list and press RES to select the unit required. Possible flowrate units are µl /h, µl/m, ml/h, ml/m. The highlighted display indicates that this parameter can be changed.
Note: If the number entered exceeds the maximum flowrate possible then the pump prompts the maximum feasible flowrate at the Time position. To continue enter a flowrate smaller than the maximum.
LSP01(04)-1A Single Channel Syringe Pump Volume setting (Release Time:2017-01-06)
In parameters setting interface, turn and press RES to highlight the volume setting line. Press RES to make the numerical value position highlighted. Enter targeted volume from the membrane keypad. Press RES to set volume unit. Turn RES to scroll through the list and press RES to select the unit required. Possible volume units are µl and ml, i.e. microliter and milliliter. The highlighted display indicates that this parameter can be changed.
Note: When the volume setting is zero, the LCD will display the flowrate setting and the pump will run until manually stopped or a stall occurs.